Product Description
| Features
-Bandwidth up to 70MHZ -Up to 14000Vp-p Differential -Safety Certified Applications
-High Voltage Floating Measurement
-Voltage Surge Measurement -Power Electronics System Design -Electronic Ballast Design Included
-Sprung Hooks Red & Black
-6 VDC Adapter -Black storage case |
|
|
Model |
PT-8020 |
|
Bandwidth(-3dB) |
DC-70MHz |
|
Attenuation Ratio |
x1000,x100 |
|
Accuracy |
±2% |
|
Input voltage range |
< 1400Vp-p for X100 (about 490V RMS or 700V DC) |
|
Permitted max input voltage |
14000V(DC+AC PEAK TO PEAK) |
|
Input Impedance |
Impedance of each side ground:10MΩ/2pF Between terminal and ground: 20MΩ/1pF |
|
Output Voltage(into 50kΩ load) |
<7V |
|
Output impedance |
50MΩ |
|
Rise time |
5ns |
|
Rejected rate on common mode |
50Hz: >-80dB; 20KHz >-60dB |
|
Power Supply |
6V DC power supply (included) 4 AA Batteries (not included) |
|
Consumption |
0.3A max |
|
Size |
186 x 84 x38 mm |
Calibration Differential Probes
The Pintech probes that we offer do not need to be Calibrated after purchasing.
At the factory they have been Mechanically zero-adjusted before shipping.
It’s less than 1mV. Mechanical zero-adjusting and will not change while using.